Follow
Video·Published May 2022Published over 4 years ago

Update Course 2021: MAGNET THERAPY FOR ESOPHAGEAL ATRESIA

With Dr. Bethany Slater & Dr. Steven Rothenberg · hosted by Dr. Todd Ponsky & Dr. Daniel von Allmen

Chapter 3 of 4 · Complications

Fistula case

Complication Case: Broncho-Esophageal Fistula and Re-Division

Cued at 11:40 · stops at 12:12 · press play

Try
Intelligent Search· scoped to esophageal atresia · not medical adviceSearch the whole library →

What the experts said

21 expert statements
The Flourish device is an FDA-approved commercially available device under humanitarian device exception from Cook Medical.
ClinicalBethany Slater4:47
The Flourish device uses compression anastomosis by causing ischemia of tissue between two magnets, which then sloughs off to create the anastomosis.
ClinicalBethany Slater4:59
The Flourish device tapers down to a 10 French coupling surface.
ClinicalBethany Slater5:40
In a retrospective study of 13 patients over 17 years using Flourish magnets, all patients achieved anastomosis but all developed strictures post-operatively.
EpidemiologicalBethany Slater7:00
Two of 13 patients in the retrospective Flourish series required surgery for recalcitrant strictures after magnet placement.
EpidemiologicalBethany Slater7:30
Flourish device eligibility requires gap length less than 4 centimeters; if lengths are greater, magnets will not attract one another.
ClinicalBethany Slater8:00
Flourish device requires either absent fistula (pure EA) or previously repaired fistula.
ClinicalBethany Slater8:15
Flourish device requires gastrostomy that can accommodate 18 French catheter (diameter of gastric portion).
ClinicalBethany Slater8:25
In the presented case, a full-term male with pure EA had gap study at 3 months showing 2.6 cm gap, and Flourish device was placed.
ClinicalSteven Rothenberg10:00
The patient was heavily sedated for 12 days during magnet treatment.
ClinicalSteven Rothenberg11:20
Four days after esophageal ends came together, patient had acute decompensation and was transferred to another hospital where large fistula between esophagus and left main stem bronchus was identified.
ClinicalSteven Rothenberg11:40
At emergency surgery, the surgeon found severe inflammation, divided the fistula, and divided the esophagus again, resulting in a 5 cm gap.
ClinicalSteven Rothenberg12:20
Subsequent workup revealed an upper pouch fistula that had not been previously diagnosed, which is a contraindication to using the Flourish device.
ClinicalSteven Rothenberg14:10
By International group classification, a gap of 4 centimeters or more is required to be considered long gap esophageal atresia.
GuidelineSteven Rothenberg17:20
Dr. Rothenberg's opinion: a patient with less than 4 cm gap should be amenable to primary anastomosis, which he would do thoracoscopically with maybe 20% stricture rate.
OpinionSteven Rothenberg17:35
Dr. Slater agrees that not everyone with 2 cm gap should get a magnet; it can be used in select patients with cardiac disease or previous operations that increase anesthetic risk.
OpinionBethany Slater18:40
Magnets have been used as staged procedure: surgically approximate ends, then use magnet just for anastomosis portion rather than internal traction sutures.
ClinicalBethany Slater19:12
Dr. Ponsky's experience: used magnets post-operatively in patient with complex airway reconstruction and median sternotomy where suture brought ends together but anastomosis was too tight, avoiding difficult reoperation.
ClinicalTodd Ponsky20:40
Dr. Von Allmen's single magnet case worked but resulted in dense stricture requiring lots of dilation and G-tube feeds for long time, though avoided reoperation.
ClinicalDaniel von Allmen21:07
Dr. Rothenberg describes potential magnet candidate: 23-week premature infant, 340 grams at birth, pure atresia, who decompensates for two weeks after each thoracoscopic attempt to approximate ends.
ClinicalSteven Rothenberg21:28
Dr. Rothenberg's opinion: if a surgeon cannot sew together a 2-3 cm gap, the patient should be sent to a center that can, as that is the best outcome for the child.
OpinionSteven Rothenberg22:20
GCMD Library — educational content from recorded physician discussions, not medical advice.