StayCurrentMD · Sacrum Ultrasonography: An Accurate Alternative to Radiography for Measuring Sacral Ratio in Infants
Article1 min read·Published Oct 2020Older

Sacrum Ultrasonography: An Accurate Alternative to Radiography for Measuring Sacral Ratio in Infants

dx.doi.org shows its articles on its own site.

Read the article on dx.doi.org ↗

Article · Oct 2020 · 1 min read

In brief

In brief

Comparative study of 129 infants demonstrating that sacrum ultrasonography achieves equivalent accuracy to radiography for measuring sacral ratio (mean difference 4.6mm), offering a radiation-free alternative for detecting sacral abnormalities in infants under 6 months.

Written by the GCMD Library team from the article.

Introduction The aim of this study was to compare the compliance of sacrum ultrasonography with radiography for the measuring of sacral ratio in infants.

Materials and Methods A total of 129 infants under the age of 6 months, who were a candidate for abdominal and/or pelvic radiographs, were assessed. Sacrum ultrasonography and radiography were performed by a single radiologist using the same device. The sacral ratio was calculated for all patients undergoing ultrasonography and radiography. Agreement between two methods was calculated by Bland-Altman's chart.

Results The mean of sacral ratio was 0.70 ± 0.11 radiographically and 0.72 ± 0.05 ultrasonographically. Based on Bland-Altman's chart, the mean difference between ultrasonography and radiology was 4.6 mm (confidence intervals of 8.18 ± 5.6).

Conclusion Sacrum ultrasonography could be safely used in the investigation of sacral ratio to detect sacrum abnormalities in infants.

Read it at the source ↗

Try
Intelligent Search· scoped to this article · not medical adviceSearch the whole library →