Pedicle subtraction osteotomy in adult spinal deformity correction: Clinical and radiographic risk factors for early instrumentation failure
pubmed.ncbi.nlm.nih.gov shows its articles on its own site.
Read the article on pubmed.ncbi.nlm.nih.gov ↗Article · Nov 2021 · 1 min read
In brief
In brief
This retrospective study of 46 PSO patients identifies high preoperative pelvic incidence, pelvic tilt, and postoperative sagittal vertical axis as significant risk factors for early instrumentation failure within 6 months. All failures occurred in 2-rod constructs via rod fracture, suggesting multiple-rod systems provide protective benefit in high-risk deformity correction.
Written by the GCMD Library team from the article.
Statements in this article
Early instrumentation failure after pedicle subtraction osteotomy occurs in 19.5% of adult spinal deformity patients within 6 months of surgery.
All early instrumentation failures in this series occurred in 2-rod constructs via rod fracture.
Patients with early instrumentation failure had significantly fewer rods than those without failure (2.00 ± 0.00 vs 2.81 ± 0.995, p = .000).
Preoperative pelvic incidence was significantly higher in the early instrumentation failure group (77.33 ± 13.23) compared to the non-failure group (p = .022).
Preoperative pelvic tilt was significantly higher in the early instrumentation failure group (37.22 ± 6.46) compared to the non-failure group (p = .012).
Postoperative sagittal vertical axis was significantly higher in the early instrumentation failure group (89.96 ± 23.85) compared to the non-failure group (p = .028).
High preoperative pelvic incidence, high preoperative pelvic tilt, and high postoperative sagittal vertical axis are significant risk factors for early instrumentation failure after pedicle subtraction osteotomy.
Multiple rod constructs are protective against early instrumentation failure and may help mitigate risk.