Clinical Implementation and Validation of Automated Slice-to-Volume Reconstruction for Fetal MRI Using a Vendor-Neutral MONAI-Based Imaging Informatics Pipeline
pubmed.ncbi.nlm.nih.gov shows its articles on its own site.
Read the article on pubmed.ncbi.nlm.nih.gov ↗Article · Aug 2026 · 1 min read
In brief
In brief
This study validates a fully automated, vendor-neutral pipeline for fetal MRI reconstruction using MONAI Deploy Express integrated with clinical PACS workflow. Among 65 examinations at a tertiary fetal center, the system achieved 90.8% acceptable brain reconstructions and demonstrated reliable operational performance for motion-corrected imaging in routine prenatal care.
Written by the GCMD Library team from the article.
Statements in this article
The automated pipeline uses MONAI Deploy Express with DICOM routing to perform slice-to-volume reconstruction for fetal MRI examinations.
The pipeline performs automated series selection and motion-corrected SVR reconstruction for fetal brain and body imaging using routine single-shot fast spin-echo inputs.
Reconstructed volumes are returned to PACS for radiologist review.
Three fellowship-trained fetal radiologists independently scored reconstruction quality using a 5-point Likert scale.
Among 67 triggered examinations, 2 represented technical workflow events and were excluded from clinical performance analysis, leaving 65 examinations for image-quality evaluation.
Mean gestational age was 28.9 weeks with a range of 19-36 weeks.
Brain SVR was rated pass-or-higher in 59 of 65 examinations, representing a 90.8% success rate.