A novel minimally invasive neurosurgical cranial fixation device for improved accuracy of intraventricular catheter placement: an experimental animal study
pubmed.ncbi.nlm.nih.gov shows its articles on its own site.
Read the article on pubmed.ncbi.nlm.nih.gov ↗Article · Dec 2024 · 1 min read
In brief
In brief
This study introduces a novel cranial fixation device designed to improve accuracy and reduce invasiveness of external ventricular drain placement in thin pediatric skulls. The device was validated through ex-vivo mechanical testing and in-vivo experiments using hydrocephalic lamb models, demonstrating enhanced fixation strength and catheter placement precision compared to traditional freehand techniques.
Written by the GCMD Library team from the article.
Statements in this article
External ventricular drain insertion is one of the most commonly performed neurosurgical procedures.
The cranial fixation device reduces reliance on freehand placement and drilling techniques for external ventricular drain insertion.
The device provides strong fixation to minimal thickness skulls.
Ex-vivo experiments included functionality bench testing and pullout force measurements for the ball mechanism and catheter fixation.
In-vivo testing was performed on one day of life 1 male control lamb and 3 hydrocephalic day of life 0 lambs.
The hydrocephalic animal model was created with in utero intracisternal injection of BioGlue in fetal lambs.
Catheter insertion trajectory was determined using MR imaging to assess the device's impact on placement accuracy.