Francesco T. Mangano · A novel minimally invasive neurosurgical cranial fixation device for improved accuracy of intraventricular catheter placement: an experimental animal study
Follow
Article1 min read·Published Dec 2024

A novel minimally invasive neurosurgical cranial fixation device for improved accuracy of intraventricular catheter placement: an experimental animal study

pubmed.ncbi.nlm.nih.gov shows its articles on its own site.

Read the article on pubmed.ncbi.nlm.nih.gov ↗

Article · Dec 2024 · 1 min read

In brief

In brief

This study introduces a novel cranial fixation device designed to improve accuracy and reduce invasiveness of external ventricular drain placement in thin pediatric skulls. The device was validated through ex-vivo mechanical testing and in-vivo experiments using hydrocephalic lamb models, demonstrating enhanced fixation strength and catheter placement precision compared to traditional freehand techniques.

Written by the GCMD Library team from the article.

Francesco T. Mangano

Background: External ventricular drain (EVD) insertion is one of the most commonly performed neurosurgical procedures. Herein, we introduce a new concept of a cranial fixation device for insertion of EVDs, that reduces reliance on freehand placement and drilling techniques and provides a simple, minimally invasive approach that provides strong fixation to minimal thickness skulls. Methods: An experimental device for catheter insertion and fixation was designed and tested in both ex-vivo and in-vivo conditions to assess accurate cannulation of the ventricle and to test the strength of fixation to the skull. The ex-vivo experiments were conducted at Ben-Gurion University of the Negev (BGU) in Be'er Sheva, Israel. These experiments included functionality bench testing and pullout force measurements for the ball mechanism and catheter fixation. For the in-vivo experiments the fixation device was initially tested at the Cincinnati Children's Hospital Medical Center (CCHMC) in Cincinnati, Ohio on one day of life 1 (DOL 1) male control lamb. Additional experiments were conducted on 3 hydrocephalic DOL 0 lambs (1 male 2 female) at the Jesús Usón Minimally Invasive Surgery Centre (JUMISC) in Caceres, Spain. The hydrocephalic animal model used for this study was created with in utero intracisternal injection of BioGlue in fetal lambs. The catheter insertion trajectory was determined using MR imaging to assess the device's impact on the placement accuracy. The fixation device was evalua

Read it at the source ↗

Try
Intelligent Search· scoped to this article · not medical adviceSearch the whole library →
Only a few other public items share this field — go deeper there →